Unique instrument advances pure tone air conduction threshold
and screening tests for hearing.
• The Model 1000 + OTO-SCREEN is a standard threshold
& screening audiometer with the option of sequential testing.
• Sequential screening eliminating fatigue factor.
• OTO-SCREEN is microprocessor controlled resulting
in accurate, consistent hearing tests.
• Hands-off screening gives the operator complete
control of testing procedure.
Industry’s newest and only PURE TONE AUDIOMETER with sequential screening
AMBCO MODEL 1000
+ OTO-SCREEN
SPECIFICATIONS
• Meets or exceeds ANSI 3:6 - 1996. ANSI 3.21 1978
(R 1997) Methods for pure-tone threshold audiometry.
• O.S.H.A. C.H.D.P. and other requirements
• Standard Mode - dB (Threshold Level) 0 to 90dB
in 5dB steps. Frequency Range: 250, 500, 1000, 2000, 3000, 4000,
6000, 8000 Hz.
• OTO-SCREEN Mode - dB (Threshold Level) 10 to 35dB
in 5dB steps Frequencies presented sequentially: 1000, 2000, 3000,
4000 Hz.
• Easy to read displays, quiet soft membrane switches
to facilitate testing.
• Portable package with padded bag, weighs 5 lbs.
5 YEAR LIMITED WARRANTY
OTO-SCREEN SEQUENTIAL TESTING:
• The Model 1000 + OTO-SCREEN is a standard threshold
& screening audiometer with the option of sequential testing.
• By the push of a button the sequential mode allows
the operator to select the screening dB (Threshold). (10-35 dB
in 5dB steps).
• Sequential screening is conducted at the school
testing frequencies, 1000, 2000, 4000 Hz or the medical screening
at 1000, 2000, 3000, 4000 Hz.
• The screening starts with a 40 dB familiarization
tone.
• A positive response by the patient with the patient
response switch initiates the sequential test tones presented
at 1000 Hz, 2000 Hz, 3000 Hz, and 4000 Hz to each ear.
• Sequential testing proceeds only after each timely
response.
• If there is no response of a false response the
fail light alerts the operator.
• Tones are presented at random to eliminate second
guessing.
• Timely responses advance to the nest frequency
until both ears are tested.
• A pass light indicates to the operator the test
was completed successfully.
• The operator is in control and can select standard
threshold or screening mode by the push of a button.